Dosing Pump Selection: A Comparison of Solenoid, Mechanical Diaphragm, and Peristaltic Pumps

KRONOS 50

A dosing pump is used not to transport a liquid, but to dispense it in a measured and repeatable manner. In water treatment, chlorine and coagulants; in boiler rooms, corrosion inhibitors; in the food industry, pH regulators; and in swimming pools, disinfectants—all are dispensed using dosing pumps. Since the expectations for accuracy are high, the choice of technology is much more critical than for transfer pumps.

In this guide, we compare three main technologies—solenoid (electromagnetic) diaphragm, motor-driven mechanical diaphragm, and peristaltic—; we discuss what accuracy means, the two issues that cause the most problems in the field (back pressure and gas leakage), and how the line should be installed. If you’re experiencing problems with an operational line, you can refer to the troubleshooting table.

Three technologies, three different operating principles

Solenoid (electromagnetic) diaphragm-type

An electromagnet pushes the diaphragm in a pulsating motion along with the piston to which it is attached; the spring pulls it back. Each pulse displaces a fixed volume; the flow rate is adjusted by the pulse frequency (strokes per minute) and stroke length.

  • Strengths: Compact, economical, wide adjustment range, ease of digital control. The most common solution for low flow rates (ranging from a few mL/hour to 80 L/hour).
  • Limitations: It is strained under high back pressure and high flow rates; stroke priming may be insufficient with viscous fluids.
  • Among the solenoid dosing pumps we offer, the Seko Tekna and Kompact series fall into this category; the Tekna ATEX option is available for sites requiring ATEX compliance.

    Motor-driven mechanical diaphragm

    The electric motor mechanically moves the diaphragm via an eccentric shaft or crankshaft. The reciprocating motion is stronger and more stable than that produced by a solenoid.

    • Strengths: High back pressure and high flow rate; suitable for continuous (24/7) operation; better priming performance with viscous liquids; high accuracy.
    • Limit: Larger, more expensive; flow rate adjustment is generally based on stroke length and motor speed.
    • Our line of mechanical dosing pumps includes the Spring, Kosmo, and Digital Elektra series; for heavier-duty applications, we offer the Invikta series.

      A higher-class variant of this family is the hydraulic diaphragm pump: the diaphragm is not actuated by a mechanical lever, but rather by oil pressurized by a piston. Since the diaphragm is subjected to pressure from both sides, it does not experience mechanical stress; this allows for very high pressures and extends the diaphragm’s service life. This is the type sought after in petrochemical and high-pressure process lines.

      Peristaltic (tube) pump

      The fluid is contained solely within the hose; the rotating rollers squeeze the hose to force the fluid through. No mechanical parts of the pump come into contact with the fluid.

      • Strengths: Excellent performance with gaseous, settling, corrosive, and viscous liquids. No valves → no place for clogging or gas buildup. Can run dry; self-priming. Material compatibility is reduced to a single component (hose).
      • Limitations: The hose is a consumable and must be replaced periodically; its pressure capacity is limited compared to diaphragm-type models.
      • Our line of peristaltic dosing pumps includes the Kronos series, featuring the PR, PM, PTS, and PRT models.

        The only factor that needs to be planned for in peristaltic pumps is hose life, and this life is determined not by a schedule but by the number of compressions. Operating the pump at a low speed (selecting a larger size and reducing the speed if necessary) directly extends hose life. Other factors that shorten service life include high pressure, high temperature, and chemicals that cause the hose to swell. On critical lines, the hose is replaced on a scheduled basis before it bursts; many models feature a leak detector that alerts the operator to a hose rupture.

        Comparison Table

        SpecificationSolenoid-diaphragmMotor-driven mechanical diaphragmPeristaltic
        Typical flow rate rangeVery low – mediumMedium – highVery low – medium
        Back pressure capacityMediumHighLow – medium
        Gas-releasing liquid (e.g., hypochlorite)Weak (degassing valve required)MediumVery good
        Viscous liquidWeakGoodVery good
        Liquid containing solids/precipitatePoorWeakVery good
        Dry runningLimitedLimitedProblem-free
        ConsumablesDiaphragm, valve ballsDiaphragm, valvesHose (regular replacement)
        Initial investmentLowHighMedium

        What does "accuracy" mean? Definitions of terms in the catalog

        Dosing pump catalogs use terms that sound similar but refer to different things. Distinguishing between them is a prerequisite for comparing two quotes. API 675, the specification for positive displacement pumps used in the petroleum, chemical, and gas industries, defines these concepts and specifies that steady-state accuracy is ±1% across the entire turndown range.

        • Steady-state accuracy: The deviation between the set flow rate and the actual flow rate. This is the key question.
        • Repeatability: The closeness of consecutive measurements taken under the same conditions. If a pump consistently delivers 5% less than the set value, its repeatability is good—but not its accuracy. This distinction is important because a repeatable deviation can be corrected through calibration.
        • Linearity: When the setting is adjusted to 50%, the flow rate actually drops by half. Proportional control does not work as expected in a pump with poor linearity.
        • Turndown (adjustment range): The highest/lowest flow rate at which accuracy is maintained. A 10:1 turndown means the pump operates with the specified accuracy even at 10% of its rated flow. A wide adjustment range listed in the catalog does not mean that accuracy is guaranteed across the entire range—you should ask specifically in which range the accuracy applies.
        • API 675 covers heavy-duty process applications; this class of pump is not required for water treatment and pool applications. However, the definitions of the terms are the same for each class and provide a common language when comparing quotes.

          Choosing the Right Flow Rate: The 20–80% Rule

          The most common mistake with dosing pumps is buying a model that’s much too large, thinking, “I might need it later.” Dosing pumps operate most accurately within 20–80% of their capacity. Running a 100-liter-per-hour pump at 5% of its capacity significantly compromises accuracy: as the stroke length shortens, valve behavior and priming become unstable.

          The right approach: Calculate your actual need and choose the model where that value falls within the pump’s mid-range.

          Example: You want to add 2 ppm of chlorine to a line with a flow rate of 50 m³/hour, and you have a 10% sodium hypochlorite solution.
          Hourly pure chlorine requirement: 50 m³/hour × 2 g/m³ = 100 g/hour.
          %For a 10% solution: 100 g ÷ 0.10 = 1,000 g/hour ≈ 1 L/hour.
          In this case, a pump with a capacity of 2–4 liters per hour is ideal; a 20-liter-per-hour model would reduce accuracy.

          The general form of the equation is as follows, and it works the same way for every chemical:

          Dosing flow rate (L/h) = [process flow rate (m³/h) × target dose (g/m³)] ÷ active ingredient concentration of the product (g/L)

          Two points throw off the calculation. First, the label percentage of a commercial solution is not the same as the active ingredient concentration; using the percentage directly as g/L for products with a density other than 1 kg/L (such as acids and caustic soda) leads to errors. Second, products like hypochlorite degrade in storage: a dosage that is correct on the first day will result in an underdose a few weeks later. Therefore, dosing is not a task that can be set once and forgotten; measurement-based monitoring automatically corrects this deviation.

          When calculating flow rate, include both the minimum and maximum requirements. The pump must meet 80% of the maximum requirement; the minimum requirement must not fall below 20%. If the difference between the two extremes exceeds the pump’s turndown ratio, a single pump will not suffice—either two pumps or a model with wider frequency/stroke control is required.

          Back pressure and siphoning: two silent faults

          The dosing pump operates on a volumetric basis; it dispenses a specific volume with each stroke. However, if the pressure balance in the system is disrupted, the dosing volume may deviate from the expected value.

          • Siphoning: If the pressure at the dosing point is lower than the level in the chemical tank, the liquid will continue to flow on its own even if the pump stops. Result: Overdosing and an empty tank. The solution is to install a back pressure valve or a siphon breaker at the dosing point.
          • Insufficient back pressure: For the same reason, the pump delivers "excess" flow at very low back pressure. The back pressure valve also stabilizes the flow rate in this situation.
          • Excessive pressure: If the discharge line becomes blocked, the positive displacement pump will continue to pump and may cause the line to burst. A safety valve must be installed on every dosing line.
          • A standard dosing line is typically set up in the following order: suction strainer and suction tube with level sensor → pump → pulse dampener (if necessary) → safety valve → back pressure valve → injection valve.

            These two components are often misunderstood. The safety valve protects the line, and the discharge must return to the chemical tank—not to the floor. The injection valve, on the other hand, is not merely a connecting piece: the spring-loaded check valve inside it prevents the process fluid from flowing back into the dosing line, and the orifice at its tip ensures that the chemical is delivered into the center of the main flow. An injection against the pipe wall causes the chemical to concentrate there and corrode the pipe.

            Suction line: a problem specific to reciprocating pumps

            The flow from the dosing pump is not continuous but pulsating: during the suction stroke, the mass of fluid in the line is suddenly accelerated. In a long, narrow suction line, this acceleration causes an additional pressure loss, and the pump cannot complete its priming cycle even with a suction flow that appears sufficient on paper. The result is a dosing volume that is not detected during calibration but is insufficient during operation.

            Practical measures are simple: keep the suction line short and wide at the pump inlet, minimize the number of elbows, position the tank above the pump if possible (to prevent flooded suction), and increase the suction diameter by one size when handling viscous fluids. For the equivalent principles in centrifugal pumps, please refer to our NPSH and cavitation guide.

            Sodium hypochlorite: a classic case of "the pump isn't drawing water"

            The most common complaint in water treatment and pool applications is that the pump stops pumping after a while during hypochlorite (bleach) dosing. The cause is not a pump malfunction: sodium hypochlorite decomposes over time and releases gas. The resulting bubbles accumulate in the valve seats of the diaphragm pump, preventing the valve from seating properly and causing the pump to lose suction (gas locking).

            Applications that prevent this:

            • Select a pump head with an automatic degassing (air release) valve; it returns the gas that accumulates during each cycle back to the tank.
            • Set up a flooded suction line: Position the pump below the tank level, and have the suction line run downward from the tank to the pump. This way, any air bubbles will return to the tank rather than entering the pump.
            • Keep the suction line short and do not leave any elevation that could create an air pocket.
            • Use up the stock quickly and keep it cool. Deterioration is accelerated by heat and light; leaving the tank in the sun will exacerbate both gas-related issues and dosing inaccuracies.
            • Choose the right material: PP is suitable for use with hypochlorite only under limited conditions; PVDF and PTFE are much safer. Among elastomers, EPDM is not as suitable, while FKM is a better choice.
            • Alternatively, switch to a peristaltic pump. Since it has no valve, gas lock is physically impossible; for this reason, peristaltic solutions have become widely used for hypochlorite dosing.
            • Material Compatibility and Calibration

              The parts of the dosing pump that come into contact with the fluid are the pump head, diaphragm, valve balls and seating surfaces, O-rings, suction/discharge hoses, and the injection valve. The weakest link in the chain determines the pump’s service life.

              ChemicalCompatible headCompatible elastomerNote
              Sodium hypochloritePVDF (preferred), PVCFKM / PTFEA degassing valve is recommended
              Sulfuric acid (diluted)PP, PVDFPTFE / FKMBehavior varies with concentration
              Hydrochloric acidPP, PVDF, PTFEPTFE / FKM316 stainless steel is not suitable
              Sodium hydroxide (20%)PP, PTFEEPDM / PTFEAvoid PVDF in hot solutions
              Polyelectrolyte / coagulantPP, PVDFEPDM / FKMViscous → peristaltic is preferable

              For detailed material specifications, please refer to our chemical compatibility guide; you can also look up the chemical you plan to transport directly in our compatibility table tool.

              Calibration: The catalog flow rate is for reference only; the actual flow rate varies depending on the line pressure and the density of the fluid. Use a calibration cylinder during commissioning: have the pump draw fluid from the cylinder, measure the volume drawn over a specific time period, and adjust the settings accordingly. If this step is skipped, the dosing error can easily reach 20%.

              There are three conditions for proper calibration: the measurement must be taken at the line’s actual operating pressure (measurements taken with the line drained are misleading), the pump must have been purged of air before the measurement, and the measurement must last at least a few minutes—calculations based on counting individual strokes will be inaccurate. Calibration is repeated when you fill the tank with a new chemical and after pump head maintenance.

              Control and measurement: switching to closed-loop dosing

              Constant-flow (open-loop) dosing with a fixed flow rate is sufficient where the load remains constant. If water quality varies, a closed-loop system is required: a probe measures the water quality, and the control panel sends a signal to the pump.

              • Proportional dosing: Dosing proportional to flow rate based on the pulse signal from the water meter.
              • Measurement-based dosing: 4–20 mA signal from a pH, ORP (redox), free chlorine, or conductivity probe.
              • The most commonly overlooked aspect of a closed-loop system is dead time: the time it takes for the chemical to travel from the injection point to the probe. If the probe is placed too far away, the controller responds too late and the system begins to oscillate—the dosage fluctuates between too much and too little. The probe should be placed at a point where the mixture is fully blended but the delay remains minimal.

                Our control and sensor product line includes the Control 40/42/65/100/102 and Control 800 panel series, as well as probes and calibration fluids. Regular calibration of the probes is the second requirement for dosing accuracy after the dosing pump: an uncalibrated probe sends incorrect commands to a properly functioning dosing pump.

                Why is pH a determining factor in chlorine dosing?

                It’s not enough to set up a closed-loop system; you need to understand why the target value is what it is. When it comes to chlorine dosing, the answer to this question lies in chemistry, and this is the source of the misunderstanding that causes the most financial losses in the field.

                Chlorine exists in two forms in water: hypochlorous acid (HOCl) and the hypochlorite ion (OCl⁻). The equilibrium between the two is determined by pH; the dissociation constant (pKa) is 7.53. This means that at pH 7.53, they are present in equal proportions. As pH decreases, HOCl becomes dominant, and as pH increases, OCl⁻ becomes dominant:

                pHHOClOCl⁻
                6.53≈ 91%≈ 9%
                7.53≈ 50%≈ 50%
                8.53≈ 9%≈ 91%

                The key point is this: HOCl is a significantly more potent disinfectant. Therefore, even if the "free chlorine" value you measure remains the same, the disinfecting power you achieve at pH 7 is not the same as that at pH 8.5. You use the same amount of chemical, but you do not get the same result.

                This is also reflected in official thresholds. According to the World Health Organization’s guidelines for drinking water, effective disinfection requires a contact time of at least 30 minutes with a free chlorine concentration of ≥ 0.5 mg/L and a pH < 8.0; residual chlorine must be maintained throughout the distribution system, and free chlorine at the point of delivery must be at least 0.2 mg/L. The guideline value should normally not exceed 5 mg/L.

                It is no coincidence that the condition pH < 8.0 appears alongside the threshold—it is a direct consequence of the equilibrium described above. The practical implication is clear: pH control is not an aid to chlorine dosing, but a prerequisite. In a system where pH is not controlled, the chlorine pump runs, chemicals are consumed, and the target residual appears to be achieved; but the disinfection capacity has silently declined.

                ORP is not a measure of chlorine concentration

                ORP (oxidation-reduction potential) is widely used and useful, but there is often confusion about what it measures: ORP measures the oxidizing potential of water, not the chlorine concentration.

                This difference is important because it is primarily the HOCl concentration that determines the potential, and that concentration depends on the pH. Conclusion: In a system where the pH is not kept constant, the same ORP value may correspond to different free chlorine levels. Therefore, an ORP-based control loop is reliable only under two conditions:

                1. If the pH is also being monitored—otherwise, the reading will drift.
                2. If it has been correlated with a known free chlorine value at a constant pH—that is, if the ORP threshold has been calibrated once using a laboratory measurement.

                An ORP threshold set without these two conditions being met is a number whose actual corresponding chlorine level is unknown.

                How is free chlorine measured?

                The determination of free and total chlorine in water is standardized by ISO 7393. The most common method used in routine testing is the DPD colorimetric method defined in ISO 7393-2: the density of the resulting red color is compared using a photometer or a calibrated color scale. In practice, the method is applicable in the range of approximately 0.03–5 mg/L.

                An annex to the standard provides a procedure for distinguishing between the forms of bound chlorine: monochloramine, dichloramine, and nitrogen trichloride. This distinction is important in plant operations: if total chlorine appears high while free chlorine is low, it means the chlorine has combined with ammonia—there is residual chlorine, but the disinfecting power is below expectations.

                Practical conclusion: The specification must state which parameter the control loop measures. The phrase “chlorine measurement” is not sufficient; it must specify whether it is free chlorine, total chlorine, or ORP. These three represent different parameters and cannot be controlled using the same setpoint.

                Chemical Storage and Safety

                The safety of the dosing line doesn't end with the pump; the tank side is just as important:

                • Secondary containment (bund). The tank is placed inside a basin capable of containing its contents in the event of a leak.
                • Keep incompatible chemicals separate. Acid and hypochlorite must not be stored in the same container: if they mix, chlorine gas is released. This is one of the most serious accidents that can occur at dosing facilities.
                • Level monitoring. The level sensor on the suction tube prevents both dry running and the unnoticed "dosing into an empty tank" situation.
                • Ventilation and eye wash. In enclosed dosing rooms, there must be ventilation and an eye wash/shower near the chemical filling station.
                • From Symptom to Cause: Rapid Troubleshooting

                  SymptomPossible causeInitial Check
                  The pump is running, but the chemical isn’t flowingGas lock, valve not seating, air leakDegassing valve, suction connection, valve balls
                  Dosage is lower than expectedCalibration is outdated, product has degraded, suction line is too long/narrowMeasurement using the calibration cylinder; product expiration date
                  Dosing is higher than expected; tank is emptying quicklySiphoningBack pressure valve, dosing point elevation
                  Dosing does not change proportionally when the flow rate is adjustedOperating outside the linearity range (setting too low)Adjust the setting to the 20–80% range; the model may be one size too small
                  The line is vibrating; connections are looseningPulsePulse damper and precharge pressure
                  Safety valve is opening frequentlyThe discharge line is clogged or the injection valve is closedInjection valve, line blockage, crystallization
                  Diaphragm tears frequentlyExcessive pressure, incompatible material, dry runningOperating pressure, material compatibility, level protection
                  The hose in the peristaltic pump bursts prematurelyHigh speed/pressure, chemical swellingReduce speed, review hose rating
                  Control panel is oscillatingProbe is too far away (dead time) or not calibratedProbe position and calibration

                  Quick Selection Summary

                  • Low flow rate, clean chemical, economical solution → solenoid-diaphragm type
                  • High flow rate, high pressure, continuous operation → motor-driven mechanical diaphragm (hydraulic diaphragm for very high pressure)
                  • Chemicals that produce gas, form precipitates, are viscous, or are corrosive → peristaltic
                  • Frequently Asked Questions

                    What flow rate (in liters per hour) should be selected for the dosing pump?

                    Your peak demand should fall within the pump’s 80% range, and your minimum demand should be above 20%. Choosing a larger pump “just in case” reduces accuracy. If the difference between the two extremes exceeds the pump’s turndown ratio, a single pump will not suffice.

                    Is a back pressure valve required?

                    It is practically essential that the pressure at the dosing point be lower than the level in the chemical tank in every case—otherwise, the liquid will flow by siphoning even if the pump stops. It also maintains a stable dosing rate even at very low back pressure. It should not be confused with a safety valve; the two serve different functions and are often found together.

                    Which dosing pump is better for hypochlorite dosing?

                    If gas discharge is an issue, a peristaltic pump offers a structural advantage: since it has no valve, gas lock is not possible. If a diaphragm pump is preferred, a head with a degassing valve, a flooded suction port, and a combination of PVDF and FKM materials are required.

                    How often should the peristaltic tube be replaced?

                    It wears out based on the number of compression cycles, not the calendar; therefore, speed, pressure, and temperature determine its service life. Operating the pump at low speed extends its service life. Hoses on critical lines are replaced on a scheduled basis before they burst, and leak detectors are used.

                    When should I repeat the calibration?

                    When commissioning the system, when filling the chemical tank with a new product, after servicing the pump head or diaphragm, and when the line pressure changes. Periodic inspections are also required for products that degrade (such as hypochlorite).

                    If you provide the chemical’s name, concentration, target flow rate, and line pressure, we can work together to determine the appropriate model and material combination—get in touch or browse our dosing pumps category.

                    Explore our measurement and control equipment, including solenoid, mechanical diaphragm, and peristaltic metering pumps.

                    Dosing Pumps Get a Quote

                    Sources and Standards

                    References for the standards and regulations cited in this guide, together with general definitions of the concepts covered. The technical values come from manufacturer data sheets and our own product data.

                    1. API Std 675 — Positive Displacement Pumps: Controlled Volume (controlled-volume dosing pumps; definitions of steady-state accuracy, linearity, and turndown) — American Petroleum Institute
                    2. Regulation (EC) No. 1935/2004 — Framework legislation on materials and articles intended to come into contact with food, EUR-Lex
                    3. Sodium hypochlorite — decomposition behavior and chemical properties — Wikipedia
                    4. Teflon (PTFE) — Chemical Resistance and Temperature Behavior — Wikipedia
                    5. Polypropylene — chemical resistance and temperature limits — Wikipedia
                    6. Elastomers — General Properties of Gasket, O-Ring, and Diaphragm Materials — Wikipedia
                    7. pH — The Foundation of Measurement and Control Applications — Wikipedia
                    8. Atlas Proses Product Data Sheets — Technical Specifications for the Seko Tekna, Kompact, Spring, Kosmo, Elektra, Invikta, and Kronos Series
                    9. Wikipedia — Hypochlorous acid: HOCl ⇌ OCl⁻ equilibrium and dissociation constant (pKa 7.53); species distribution that varies with pH
                    10. World Health Organization (WHO) — Chlorine in Drinking Water, Chemical Information Note from the Guidelines for Drinking Water Quality (≥0.5 mg/L free chlorine, ≥30 min contact time, pH < 8.0; ≥0.2 mg/L at the point of delivery)
                    11. ISO 7393-2:2017 — Water quality: Determination of free chlorine and total chlorine, Part 2: Colorimetric method using N,N-dialkyl-1,4-phenylenediamine (DPD method; differentiation of bound chlorine forms)